beautypg.com

Cooper Instruments & Systems DFI 1650 Multi-Channel Digital Force Indicator User Manual

Page 33

background image

CF 66

27

Rev. C 2/05

FREQ.RESPONSE Menu Item
This sets the frequency response, step response, and resolution of the channel according to the table below. The
tracking value, peak/valley detector values and analog output on this channel are all affected. The default value of
016. HERTZ is suitable for most applications.

Frequency

Response

Step Response

(ms) (typical)

Resolution (counts) (not including

min. 10% overrange/underrange

capability)

002. HERTZ/FAST

40

±50000

002. HERTZ

440

±50000

008. HERTZ

110

±25000

016. HERTZ

55

±20000

032. HERTZ

28

±10000

050. HERTZ

16

±5000

100. HERTZ

8

±5000

250. HERTZ

3

±2000

500. HERTZ

2

±2000

800. HERTZ

2

±2000


POWER-ON SOURCE Menu Item
This menu selects which value is displayed by the channel when first entering the RUN mode.

The choices are:

• “TRACK” means the live tracking value of the channel

• “PEAK” means the highest value of the channel.

• “VALLEY” means the lowest value of the channel.

• “TRACK/PEAK” means the live tracking value and the highest value of the channel (only available on DFI

SC3004)

• “TRACK/VALLEY” means the live tracking value and the lowest value of the channel (only available on DFI

SC3004)

• “PEAK/VALLEY” mans the highest and lowest values of the channel. (only available on the DFI SC3004)

10.5.2 DISPLAY SETUP Sub-Menu

This menu controls how data values are displayed by the channel and transmitted via serial communications.

DISPLAY.DIGITS Menu Item

Selects the number of digits displayed by the channel. The choices are:

• “5 DIGIT-BIPOLAR” displays both positive and negative numbers with five full digits (±999999).

• “6 DIGIT-UNIPOLAR” displays positive numbers with six full digits (999999) and negative numbers with five full

digits (-99999).

• “7 DIGIT UNIPOLAR” displays positive number with seven full digits (9999999) and negative numbers with six

digits (-999999).


This menu item may be automatically updated by a transducer’s Signature Module.

DISPLAY.DECPT Menu Item
Selects the decimal point location on the channel’s display and serial communications output. Use the [UP] and
[DOWN] buttons to move the decimal point to the right and left.

This menu item may be automatically updated by a transducer’s Signature Module.

DISPLAY.COUNT-BY Menu Item
Determines by what value the display increments or decrements. This will make the display less sensitive to a
noisy signal. The choices are:

• “00001”

• “00002”

• “00005”