BioLogic LP System
Power
115 V ~ 5.3 A
230 V ~ 3.5 A
50 - 60 Hz
Operating temperature
2° to 40°C,
≤
95% humidity
Construction Material
Polypropylene and other solvent resistant plastics
Flow rate range
0.05 to 20 ml/min (depending on tubing diameter)
(per channel)
Pump head speed
25 rpm (max.)
Tubing diameter
0.4 mm (ID) to 3.2 mm (ID), maximum 1 mm wall thickness
Speed adjustment
.01 ml/min
Speed stability
1% full scale
Counterpressure (max.)
30 psi (2 kg/cm2 or bars)
UV Monitor
Mercury/Phosphor Lamp
Wavelength: 280 and 254 nm (both filters supplied)
Auto Zero Feature
Eleven fixed ranges, 2.0 to 0.001 AU Full Scale
User set range 2.0 to 0.001 AU Full Scale
Cuvette volume: 80 µl
Path length: 2 mm
Illuminated volume: 3 µl
Full Scale output to chart recorder: 1 volt
Non Scaled Integrator output: 1 volt @ 2 AU
Conductivity Monitor
AC excitation
Nine fixed ranges, 500 to 0.5 mS/cm Full Scale
User Set minimum and maximum values from 500 to 0.5 mS/cm
Flow Cell swept volume: 8 µl
Accuracy: Within 2% of full scale
Full Scale output to chart recorder: 1 volt
Proportioning Accuracy: Within 1 % of full scale (3% to 97% B)
Gradient Linearity: Within 3 % of full scale (3% to 97% B)