Table 24. (continued) run mode: starting a run – Bio-Rad BioLogic LP System User Manual
Page 84

Table 24. (continued)
Run Mode: Starting a Run
Reference
Description
Multiruns: This allows the method to be run the method from 1
to 999 times. Enter the number of runs desired using the num-
ber keys. The OK softkey returns to the Countdown screen
with the number of runs entered. The Cancel softkey returns to
the Countdown screen with the original number of runs prior to
selecting Multiruns.
Note: The Multirun feature requires a fraction collector capacity
of the number of tubes for each run times the number of runs.
For example, if 25 fractions are to be collected in each run and
Multirun is set to 3 then 25 x 3 = 75 fractions will be collected.
Delay: This allows the delay setting to be entered with a value
between 0 and the smallest fraction size programmed within the
method. Enter the delay setting desired using the number keys.
The OK softkey returns to the Countdown screen with the delay
setting entered. The Cancel softkey returns to the Countdown
screen with the original delay setting prior to selecting Delay.
Note: For a description of the Delay feature see section 5.4.1.2
below.
77
System Operation
R1a
Method: <
Volume: 0.51 ml
Number of Runs: 1
CANCEL
OK
(returns
to R1)
R1b
Method: <
Volume: 0.51 ml
Delay: 1 ml
CANCEL
OK
(returns
to R1)