Table 10. (continued)
Manual Mode Operation: Conductivity Monitor
Reference
Description
Min/Max: Allows you to set a range, from 0.00 mS to 999.99
mS, not selectable under Set Range’s incremental values. To
use this function, run buffer A through the system until the con-
ductivity reading stabilizes, and note the conductivity reading.
Then run buffer B through the system and note its conductivity
reading. In the conductivity Min/Max screen, enter the noted A
and B values as the minimum and maximum.
45
System Operation
Method: <
Manual Operation
Conductivity Monitor: 0.0111 mS
Range: Min: 0.00 mS
Max: 5.00 mS
(set min)
SET MAX
CANCEL
OK
M3
MU
MU
(Set the Minimum and
Maximum values for
the range of detection)
Method: <
Manual Operation
Conductivity Monitor:
0.00 mS
Range:
500.0 mSFS
SET RANGE
CALIBRATE
MIN/MAX
Conductivity Cell Constant:
40.0000
(constant)
SOLUTION
CANCEL
OK
Enter the Conductivity of the solution
now in the Conductivity Flow Cell:
0.00 mS/cm
CONSTANT
(solution)
CANCEL
OK
(return to MC)
(return to previous screen)
(return to MC)
(return to MC)
Method: <
Manual Operation
Conductivity Monitor:
0.0111 mS
Range: 500.0 mSFS
INCREASE
DECREASE
CANCEL
OK
Method: <
Manual Operation
Conductivity Monitor:
0.0111 mS
Range: Min: 0.00 mS
Max: 5.00 mS
(set min)
SET MAX
CANCEL
OK
press to increment/
decrement the range value
by approximately 50% each time
the button is pressed
(return to MC)
(Set the
maximum value)
(set minimum
value is
selected)
MC
M2a
MC
MANUAL MODE: CONDUCTIVITY
SET CONDUCTIVITY RANGE
CALIBRATE FOR CONDUCTIVITY
M1
M2
SET MIN/MAX VALUES FOR CONDUCTIVITY
M3
M1
M3
M2
LEGEND
Figure 19. Manual Operation of the Conductivity Monitor
(Screen displays are keyed to the discussion in Table 10, Manual Mode Operation : Conductivity Monitor)