B | Troubleshooting
218 | ChemiDoc MP Imaging System with Image Lab Software
Printout does not
look like the monitor
image
The monitor settings are
wrong.
The printer settings are
wrong.
See your monitor manual for the
appropriate settings.
See your printer manual for the
appropriate settings.
Light leakage into the
darkroom
The lens body is not
seated properly against
the gasket on the hood’s
adapter plate.
Loosen the thumbscrew and seat the
lens properly against the gasket on
the hood’s adapter plate.
Unable to focus on
the sample using
white light
transilluminator or
conversion screen
Focus is not calibrated
for samples using this
light source.
Select Edit > Instrument Setup to
recalibrate the focus for use with this
accessory.
Lens limits seem
artificially restricted
The camera lens is not
seated properly on the
lens mounting plate.
Re-seat the camera on the lens
mounting plate.
Problem
Possible Cause
Solution