Pr.24, L, f, s – Kenco Engineering SmartSonic Transmitter (Integral) User Manual
Page 8
Page 8
Five Initial Programs
While the
Pr.02 Program is normally used to set the Tank Height, there is a special code that is used to enter this Initial
Set-up Program. The following directions show you how to get into the Initial Set-up Program:
Press / Action
Display
Description
Connect SmartSonic to power supply
Temporary display while SmartSonic takes a
reading
After a brief pause
For Example:
Measurement reading; Normal Mode
and
simultaneously
Enter the Program Menu
One Time
Used to search for the required menu selection.
For this step, find Program
Pr.02
Indicates the measurement units: meters
(shown) or feet (H F).
For Example:
Displays last saved Tank Height or Maximum
Range
and
Change the display to read 00.18
Now use the “Back” button to go to
Pr. 24
Initial Set-up Program Pr.24 (Application Type)
This program allows you to set the transmitter to either Level (L) or Flow (F) Measurement. DO NOT select “S”, as this
function is not activated at this time.
Press / Action
Display
Description
Pr.24
Pr.24
Pr.24
Pr.24
Program #24
or
L, F, S
L, F, S
L, F, S
L, F, S
Cycle between Level (L), Open Channel Flow
(F). DO NOT select “S”
Accepting the new “Application Type” Value
After a brief time
Pr.24
Pr.24
Pr.24
Pr.24