Each reading (scan distance) taken using the

Pr.03 program is stored as an interfering signal, until a reading is achieved

that indicates the real echo. Prior to running this program, you must verify the distance to the bottom of the tank (i.e. stick
the tank). If six interfering signals are already stored, any additional interfering signals will not be stored. If this occurs,
the process fluid level in the tank must not go below this point. The reading of the actual target may not be exact. For
example a target at 6 feet may give a reading in the range of 5.95 to 6.05.

NOTE: THE DISPLAY IN Pr.03 IS IN DISTANCE (FROM THE SENSOR), NOT LEVEL OR FLOW.

Press / Action

Display

Description

and

simultaneously

Enter the Program Menu




Two Times

Cycle to Program #3

Enter

Pr.03.Use BACK or NEXT buttons to

cycle between Search (SrCH) – to search for
Interference Signals; or Clear (CLr) – to clear
all existing Interference Signals




Displayed for approximately 3 seconds. Goes
to Pr.03 if you selected Clear. Goes to the next
step if you selected Search

Echo Amplitude

Interferences

Stored in

Memory

(up to 6)

Distance

Real Echo