Kenco Engineering SmartSonic Transmitter (Integral) User Manual
Page 15

Page 15
Each reading (scan distance) taken using the
Pr.03 program is stored as an interfering signal, until a reading is achieved
that indicates the real echo. Prior to running this program, you must verify the distance to the bottom of the tank (i.e. stick
the tank). If six interfering signals are already stored, any additional interfering signals will not be stored. If this occurs,
the process fluid level in the tank must not go below this point. The reading of the actual target may not be exact. For
example a target at 6 feet may give a reading in the range of 5.95 to 6.05.
NOTE: THE DISPLAY IN Pr.03 IS IN DISTANCE (FROM THE SENSOR), NOT LEVEL OR FLOW.
Press / Action
Display
Description
and
simultaneously
Enter the Program Menu
Two Times
Cycle to Program #3
Enter
Pr.03.Use BACK or NEXT buttons to
cycle between Search (SrCH) – to search for
Interference Signals; or Clear (CLr) – to clear
all existing Interference Signals
Displayed for approximately 3 seconds. Goes
to Pr.03 if you selected Clear. Goes to the next
step if you selected Search
Echo Amplitude
Interferences
Stored in
Memory
(up to 6)
Distance
Real Echo