11 connection to 761 compact ic or 790 personal ic – Metrohm 838 Advanced Sample Processor Installation Instructions User Manual
Page 59

3.5 Connection to Metrohm IC systems
Metrohm 838 Advanced Sample Processor, Installation
53
Injektion valve in FILL position
Injektion valve in INJECT position
Waste
Sample Waste
Sample
Column
Eluent Column
Eluent
838-specific settings in the «IC Net 2.3» program
Program for the 838 Advanced Sample Processor:
001 Move
Sample
002 Lift
work
003 Valve
Fill
004 Pump
240
s
005 Valve
Inject
006 Start
DA
− Move needle to sample position
− Move lift with needle to working height
− Switch over injection valve on 838 to "Fill"
− Fill sample loop with sample
− Switch over injection valve on 838 to "Inject"
− Start data recording
3.5.11 Connection to 761 Compact IC or 790 Personal IC
The compact 761 Compact IC is also controlled from the PC. The 838
Advanced Sample Processor is controlled via an RS232 connection. If
necessary, Compact IC and Sample Processor may also be synchro-
nised via a remote connection.
838
761
PC
PC
Figure 39 Interconnection with the 761 Compact IC via PC (RS232 only)
6.2134.100
6.2134.100