3.3 Fitting the accessories
838 Advanced Sample Processor, Installation
17
3.3.6 Fitting the ultrafiltration cell
This section describes initial fitting of the ultrafiltration cell including fil-
tration membrane. Please note Chapter 6.1 when assessing the need to
exchange the membrane and when selecting membranes. The proce-
dure for exchanging a filtration membrane is also described in full in
Chapter 4.2.2.
Figure 17 Fitting the ultrafiltration cell
28
Ultra filtration cell (6.2729.110)
33 Outlet for filtrate
29
Upper section of the filtration
cell
34
Filtration membrane
(6.2714.020)
30 Sealing ring (E.301.0111)
35
Inlet for sample solution
31
Lower section of the filtration
cell
36 Outlet for sample solution
32
Screws (V.022.6030) including
washer (4.754.4090)
1 Preparing the ultrafiltration cell
• Remove the ultrafiltration cell 28 from its packaging and remove the
three screwed-on blind stoppers.
• Using the 6.2621.070 Allen key (5 mm), fully undo the 5 screws 32,
disconnect the upper section 29 from the lower section 31 and re-
move the sealing ring 30.
• Thoroughly rinse off the sealing ring, lower section and upper sec-
tion of the ultrafiltration cell with ultra-pure water.
Only ultra-pure water or ethanol may be used to clean the ultrafiltration
cell. Using organic solvents (e.g. acetone) will damage the Plexiglass
cell!
29
31
30
32
28
33
34
35
36