Metrohm 838 Advanced Sample Processor Installation Instructions User Manual
Page 53

3.5 Connection to Metrohm IC systems
Metrohm 838 Advanced Sample Processor, Installation
47
3.5.8 MIC-3 modular anionic and cationic system with chemical suppres-
sion
This system requires the model 2.838.0010 of the 838 Advanced
Sample Processor.
The combined anionic/cationic IC system can operate either with
chemical suppression or with electronic suppression. The 838 Ad-
vanced Sample Processor is connected to the System MIC-3 with two
819 IC detectors, 820 IC Separation Centre (2 injectors), two 818 IC
pumps, 833 IC Suppressor Module and 830 IC Interface in accordance
with Figure 36 using the 6.2125.120 remote adapter, the 6.2125.090
remote cable and the 6.2134.100 RS232 cable.
838
819
819
830
820
818
PC
6.2134.100
6.2125.090
6.2134.100
6.2141.190
833 MSM
6.2125.120
818
6.2125.090
6.2134.090
6.2143.210
6.2115.070
6.2128.180
6.2134.040
6.2141.120
6.2134.130
Figure 36 MIC-3 interconnection with modular anionic and cationic system
with chemical suppression
838-specific settings in the «IC Net program 2.3»
Configuration of additional remote lines for the 838 Advanced Sample Processor:
Fill A 2
***0*010***
Inject A 2 ***1*000***
Reset 2
***0*000***
ZERO 2
***0*011***
Program for the 838 Advanced Sample Processor:
001 Move
sample
002 Lift
work
003 Ctrl
ZERO
1
004 Ctrl FILL A 1
005 Ctrl STEP MSM 833
006 Pump 120 s
007 Ctrl INJECT A 1
008 Move
Sample+1
009 Lift
work
− Move needle to sample position
− Move lift with needle to working height
− Trigger autozero on 819/1 IC Detector
− Switch over injection valve 1 on 820 to "Fill"
− Trigger step in the Suppressor Unit
− Fill sample loop for 120 s with sample
− Switch over injection valve 1 on 820 to "Inject"
− Move needle to next sample
− Move lift with needle to working height