beautypg.com

Figure 33, Install pump tubing connection without filter – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 68

background image

3.19 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

58

■■■■■■■■

850 Professional IC Anion MSM-HC LP Gradient

1

2

3

Figure 33

Install pump tubing connection without filter

1

Union nut

2

Adapter

3

Tubing olive

Slide union nut (33-1) onto the pump tubing.

Select a suitable adapter (33-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 2,
page 58)
.

Place the tubing olive (33-3) onto the pump tubing.

Screw the union nut (33-1) onto the tubing olive (33-3).

4 Inserting the pump tubing

Press the contact pressure lever all the way down.

Place the pump tubing in the tubing cartridge. The stoppers
(31-3) must snap into the corresponding holders of the tubing
cartridge.

5 Inserting the tubing cartridge

Hang the tubing cartridge in the mounting pin and press in the
cartridge holder until the snap-action lever snaps in.

6 Connecting the capillaries

Screw the respective capillaries tightly to the two tubing olives
with PEEK pressure screws (31-1).

Table 2

Pump tubings and suitable adapters

Pump tubing

Adapter

6.1826.020 (blue/blue)

6.1826.310 (orange/green)

6.1826.320 (orange/yellow)

6.1826.330 (orange/white)