3 Installation
850 Professional IC Anion MSM-HC LP Gradient
■■■■■■■■
13
■
Place the detector in the instrument and connect it (see manual
of the detector).
■
Remove transport locking screws (see "Removing transport lock-
ing screws", page 23).
■
Connect the leak sensor (see "Connecting the leak sensor ", page
23).
■
Connect drainage tubings (see "Installing drainage tubings",
page 26).
3 Installing the eluent path
■
Equip the eluent aspiration tubing and connect with the eluent
bottle (see Chapter 3.9.1, page 29).
■
Instead of the column, connect the 6.2744.040 coupling with a
6.2744.070 short pressure screw at the end of the pre-installed
column inlet capillary.
■
Use a 6.2744.070 pressure screw to connect the capillary of the
MSM-HC labeled with in to the other end of the 6.2744.040 cou-
pling.
■
Connect the detector inlet capillary and the capillary of the
MSM-HC labeled with out to one another using one 6.2744.040
coupling and two 6.2744.070 short pressure screws (see manual
of the detector).
4 Installing the sample path
Note
The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.
■
Guide the 6.1803.040 sample aspiration capillary connected to
the sample input of the injection valve out of the instrument
through a capillary feed-through and connect it with the Sample
Processor or
■
If the sample degasser is used connect the sample aspiration
capillary with a 6.2744.090 long pressure screw to the input of
the sample degasser (see Chapter 3.15, page 45).
Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through and connect it to the Sample Processor.