beautypg.com

9 eluent, 1 connecting eluent bottle, Eluent – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 39: Connecting eluent bottle

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

29

3.9

Eluent

3.9.1

Connecting eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing (11-1).

The eluent aspiration tubing is connected to the eluent degasser (see
Chapter 3.10, page 33)
. The tubing must be threaded through a suita-
ble capillary feed-through (see Chapter 3.8, page 26) of the instrument
before the other end can be equipped.

You will require the parts from the following accessories for equipping the
eluent aspiration tubing:

6.1602.160 Eluent bottle attachment GL 45

6.2744.210 tubing adapter for aspiration filter

6.2821.090 aspiration filter

To equip the eluent aspiration tubing proceed as follows:

Assembling eluent aspiration tubing

1 Guide the free end of the eluent aspiration tubing (11-1) out of the

instrument through a suitable capillary feed-through.

2 Installing eluent bottle attachment 6.1602.160

Slide tubing nipple (11-2) and O-ring (11-3) onto the eluent aspi-
ration tubing (11-1).

Push eluent aspiration tubing (11-1) through the bottle attach-
ment (11-4) and screw tight.