3 Installation
850 Professional IC Anion MSM-HC LP Gradient
■■■■■■■■
31
4 Install tubing weighting and aspiration filter
1
2
3
4
5
Figure 13
Install tubing weighting and aspiration filter
1
Eluent aspiration tubing 6.1834.080
2
Eluent bottle attachment 6.1602.160
3
Tubing weighting
From accessories set 6.2744.210.
4
Clamping screw
From accessories set 6.2744.210.
5
Aspiration filter 6.2821.090
With filter holder from accessories set
6.2744.210.
■
Slide the tubing weighting (13-3) onto the eluent aspiration tub-
ing (13-1).
■
Slide the clamping screw (13-4) onto the eluent aspiration tubing
(13-1).
■
Insert the eluent aspiration tubing (13-1) into the aspiration filter
(13-5). The end of the tubing should approximately reach to the
center of the aspiration filter.
■
Screw together clamping screw (13-4) and filter holder (12-1).
Figure 14
Eluent aspiration tubing fully equipped.
5 Mounting eluent aspiration tubing to the eluent bottle
■
Insert the eluent aspiration tubing into the eluent bottle (15-10).