3.18 High Capacity Metrohm Suppressor Module (MSM-HC)

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

850 Professional IC Anion MSM-HC LP Gradient

3.18

High Capacity Metrohm Suppressor Module
(MSM-HC)

3.18.1

General information on the MSM-HC

The MSM-HC is used for chemical suppression during anion analysis if elu-
ents with a high ionic concentration are used or if the concentration is
being significantly increased in the course of a gradient application. It con-
sists of 3 suppressor units in total, which are used for suppression, regen-
erated with sulfuric acid or rinsed with ultra pure water in rotation.

Suppression reaction in the MSM-HC
When using a carbonate eluent, the following reaction (amongst others)
occurs in the MSM-HC:

R-SO

3

-

H

+

+ NaHCO

3

/Na

2

CO

3

➙ R-SO

3

-

Na

+

+ H

2

O + CO

2

3.18.2

Connecting the MSM-HC

Caution

To protect the MSM-HC against foreign particles or bacterial growth, a
6.2744.180 pump tubing connection with filter must be mounted
between the peristaltic pump and the inlet capillaries of the MSM-HC.

The three inputs and outputs of the suppressor units numbered with 1, 2
and 3 on the MSM-HC each have 2 fixed mounted PTFE capillaries (see
Figure 29, page 53)
.