beautypg.com

14 pulsation damper, Figure 23, Connecting inline filter – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 54

background image

3.14 Pulsation damper

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

850 Professional IC Anion MSM-HC LP Gradient

2

3

2

4

1

Figure 23

Connecting inline filter

1

Connection capillary
Connects the purge valve with the inline fil-
ter

2

PEEK pressure screws, short
6.2744.070

3

Inline filter 6.2821.120
Protects against particles.

4

Connection capillary
Connects the inline filter with the pulsation
damper.

1 Screw on the connection capillary running from the purge valve with

a 6.2744.070 pressure screw to the input side of the inline filters.

2 Screw on the connection capillary running to the pulsation damper

with a 6.2744.070 pressure screw to the output side of the inline fil-
ter.

3.14

Pulsation damper

Note

The pulsation damper is already installed in the newly delivered instru-
ment.

Caution

The pulsation damper is maintenance-free and may not be opened.

The pulsation damper protects the separation column from damage
caused by pressure fluctuations when switching the injection valve, and
reduces interfering pulsations during highly sensitive measurements. In
order to ensure these functionalities, it must be connected between the
high pressure pump (see Chapter 3.12, page 39) and injection valve (see
Chapter 3.16, page 47)
.

The pulsation damper can be operated in both directions.