beautypg.com

7 technical specifications, 1 reference conditions, 2 device – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 111: 3 leak sensor, 4 ambient conditions, Reference conditions, Device, Leak sensor, Ambient conditions

background image

■■■■■■■■■■■■■■■■■■■■■■

7 Technical specifications

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

101

7 Technical specifications

7.1

Reference conditions

The technical specifications listed in this chapter refers to the following
reference conditions:

Ambient tempera-
ture

+25 °C (±3 °C)

Instrument status

> 40 minutes in operation (equilibrated)

7.2

Device

IC system

Metal-free IC system

Compact system with modular design

Up to two complete chromatographic systems in one housing

Material

Painted polyurethane hard foam without CFCs, fire class V0

Operating pres-
sure range

0…50 MPa (500 bar) high pressure pump

0…35 MPa (350 bar) standard-PEEK system

Intelligent com-
ponents

iPump, iDetector, iColumn, MagIC Net

7.3

Leak sensor

Type

Electronic, no calibration necessary

7.4

Ambient conditions

Operation

Ambient tem-
perature

+5…+45 °C

Humidity

20…80 % relative humidity

Storage

Ambient tem-
perature

–20…+70 °C

Transport

Ambient tem-
perature

–40…+70 °C