beautypg.com

3 deaerating the high pressure pump, Deaerating the high pressure pump, Figure 21 – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 51: Mixing coil for low pressure gradients connection

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

41

1

2

3

Figure 21

Mixing coil for low pressure gradients connection

1

Inlet valve holder

2

PEEK pressure screw, short 6.2744.070

3

Mixing coil for low pressure gradients
6.2758.020

3.12.3

Deaerating the high pressure pump

The high pressure pump will only operate perfectly if the pump head con-
tains no more air bubbles. Therefore it must be deaerated during initial
start-up and after every change of eluent.

Caution

The high pressure pump must not be deaerated before the initial start-
up (see Chapter 4.1, page 64).

Deaerate the high pressure pump as follows (see Figure 22, page 42):

Deaerating the high pressure pump

The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.