5 Operation and maintenance

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

93

Order
number

Name

Material

Inner
diameter

Use

6.1826.330

Pump tubing LFL
(orange/white), 3-
stopper

PVC (Tygon)

0.64 mm

No special applications.

6.1826.340

Pump tubing LFL
(black/black), 3-stop-
per

PVC (Tygon)

0.76 mm

For the sample solution in
inline dialysis.

6.1826.360

Pump tubing LFL
(white/white), 3-stop-
per

PVC (Tygon)

1.02 mm

For sample transfer.

6.1826.380

Pump tubing LFL
(gray/gray), 3-stopper

PVC (Tygon)

1.25 mm

For inline sample dilution.

6.1826.390

Pump tubing LFL (yel-
low/yellow), 3-stop-
per

PVC (Tygon)

1.37 mm

For the sample solution in
inline ultrafiltration.

5.12.2.2

Pump tubing connection with filter

The 6.2821.130 filters (45-2) should be changed every 3 months, more
frequently at higher backpressure.

1

2

3

Figure 45

Pump tubing connection – Changing the filter

1

Tubing olive

2

Filter 6.2821.130
Packaging contains 10 items.

3

Filter housing

Replacing the filter

1 Unscrewing filter screw

Screw the filter screw (45-3) out of the tubing olive (45-1) with
the aid of two 6.2621.000 adjustable wrenches.

2 Replacing the filter

Remove the old filter (45-2) with tweezers.

Place the new filter (45-2) flat in the tubing olive (45-1) with
tweezers.