6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
98
■■■■■■■■
850 Professional IC Anion MSM-HC LP Gradient
Problem
Cause
Remedy
Sample – Sample loop not
(completely) filled.
Prolong the sample transfer time.
Sample – Gas bubbles in
the sample.
Use the sample degasser (see Chapter 3.15,
page 45) (if available).
Peristaltic pump –
insufficient or no
delivery rate
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure (see "Set flow
rate", page 59).
Peristaltic pump – filter
blocked.
Replace the filter (see Chapter 5.12.2.2, page
93).
Peristaltic pump – pump
tubing defective.
Replace pump tubing (see Chapter 5.12.2.1,
page 92).
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves (see "Cleaning the inlet
valve and outlet valve", page 76).
Eluent – Leak in the eluent
path.
Check the eluent path
Eluent – Blockage in the
eluent path.
Check the eluent path
High pressure pump –
defective piston seals.
Replace piston seals (see "Replacing the piston
seal", page 73).
Pulsation damper not con-
nected.
Connect the pulsation damper (see Chapter
3.14, Page 44).
Pulsation damper not con-
nected.
Connect pulsation damper (see Chapter 3.14,
page 44).
Background conduc-
tivity too high
MSM-HC – not connected.
Connecting the MSM-HC (see Chapter 3.18,
page 52).
Incorrect eluent.
Change eluent (see Chapter 5.4.2.3, page 70).
MSM-HC – regeneration or
rinsing solution flow prob-
lems.
Check the flow of the regeneration and rinsing
solution (see Chapter 3.18.2, page 52).
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.