background image

■■■■■■■■■■■■■■■■■■■■■■

10 Appendix

863 Compact Autosampler

■■■■■■■■

61

Outputs

Open Collector

tp

t

p

> 200 ms

active = low, inactive = high

I

C

= 20 mA, V

CEO

= 40 V

+5 V: maximum load = 20 mA

10.1.2

Status diagram of the remote interface

[START]

Output 0

Output 1

Output 2

Output 3

Output 4

Ready

Error

Start/Continue

Stop

Input 0

Input 1

[STOP]

Figure 18

Remote status diagram

10.2

USB devices

Note

USB peripheral devices that are to be connected directly must support
the USB 1.0/1.1 (Full Speed) or USB 2.0 (High Speed) standard. The
maximum data transfer rate is however in any case 12 MBit/s.

Keyboards, PC mice and barcode readers are so-called HID devices
(Human Interface Device) and can be connected via a USB hub only.

Printers should also be connected via a USB hub. Depending on the
manufacturer or printer type a direct connection is however possible.