4 Automation sequences
863 Compact Autosampler
■■■■■■■■
27
■
Wait for the pump time to be completed
■
Switch off the peristaltic pump
■
Wait for the stepping pulse
4.5
VA 797 Remote start
This method template is suitable for sample transfer to a 797 VA Compu-
trace. The actual pump time is determined by the control software.
The sample series is started on the 797 VA Computrace (Remote Start).
Preparing and starting the sample series
1 Press [START].
2 Under Number of samples, enter the number of samples or the
maximum value of 99.
3 Under Next sample pos., enter the rack position of the first sample
to be processed.
4 Press [BACK].
5 Start the sample series in the VA Computrace software.
Note
The method run will be automatically restarted for every sample via
remote connection (Autostart). A stepping pulse of the connected
instrument is expected at the end of the method run. The 863 Compact
Autosampler must be stopped manually by means of the button
[STOP] after the processing of the last sample.
The individual steps of the method:
■
Move to the sample
■
Lower the lift to the work position
■
Wait for the stepping pulse
■
Switch on the peristaltic pump for the sample transfer
■
Wait for the minimum pump time to be completed
■
Wait for the stepping pulse
■
Switch off the peristaltic pump
■
2 seconds waiting time
■
Wait for the stepping pulse