4 configuring external devices, Configuring external devices, Figure 16 – Metrohm 863 Compact Autosampler User Manual
Page 59: Adjusting the needle

■■■■■■■■■■■■■■■■■■■■■■
6 System settings
863 Compact Autosampler
■■■■■■■■
49
3
2
Figure 16
Adjusting the needle
3 Adjust the needle holder
Turn the needle holder a few millimeters out of the lift head (turn
counterclockwise).
Now you can redefine the working height of the lift (see previous
section), and precisely set the needle position by turning the needle
holder.
4 Tighten the nut
Using the wrench, fasten the nut again.
5 Close the safety shield
Tip down the safety shield and fix it with the fastening screws.
6.4
Configuring external devices
Menu
▶ System ▶ External devices
Printer
If a printer is connected, then the printer type needs to be defined here in
order for the reports to be printed out correctly.
The printers that have the designation ESC-POS are so-called POS printers
(point-of-sale printers), i.e. they print on continuous paper.
Selection
Citizen (ESC-POS) | Custom (ESC-POS) | Epson |
Epson (ESC-POS) | HP DeskJet | HP LaserJet |
Seiko (ESC-POS)
Default value
HP DeskJet
Keyboard layout
A commercially available USB keyboard can be connected to make it eas-
ier to enter text and numbers. Specify the country-specific keyboard layout
here.