Figure 10, Connecting usb devices – Metrohm 863 Compact Autosampler User Manual
Page 29

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
863 Compact Autosampler
■■■■■■■■
19
6.2151.100
Figure 10
Connecting USB devices
Caution
Switch the instrument off before connecting or disconnecting a USB
device or a USB stick.
The 863 Compact Autosampler can only recognize the device immedi-
ately after switching on.
The following devices can be operated directly on the USB connector
with the 6.2151.100 adapter:
■
USB sticks (for the backup or storing of methods)
■
6.2147.000 numerical USB keypad
■
USB hub (with or without an own power supply)
The 6.2147.000 numerical USB keypad serves for comfortable numer-
ical input and for navigating in the dialog. In addition, it provides two USB
connectors. Connect additional USB devices to the keypad.
Note
Most of the USB devices need a so-called hub in order to work cor-
rectly.
A USB hub is a distributor to which several USB devices can be connec-
ted. USB hubs are available in specialty stores in a number of different
models.
The USB (OTG) connector of the 863 Compact Autosampler has no
such hub. The 6.2147.000 numerical USB keypad has a USB hub and
two USB connectors.
The following devices can only be connected to a 6.2147.000
numerical keypad or to a USB hub:
■
Printer (with USB connector, use the 6.2151.020 connecting cable)