beautypg.com

6 va 797 manual start, Va 797 manual start – Metrohm 863 Compact Autosampler User Manual

Page 38

background image

4.6 VA 797 Manual start

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

863 Compact Autosampler

4.6

VA 797 Manual start

This method template is suitable for sample transfer to a 797 VA Compu-
trace. The actual pump time is determined by the control software.

Preparing and starting the sample series

1 Press [START].

2 Under Number of samples, enter the number of samples or the

maximum value of 99.

3 Under Next sample pos., enter the rack position of the first sample

to be processed.

4 Press [START].

Note

The method run will be automatically restarted for every sample via
remote connection (Autostart). A stepping pulse of the connected
instrument is expected at the beginning of the method run. The 863
Compact Autosampler may need to be stopped manually by means of
the button [STOP] after the processing of the last sample.

The individual steps of the method:

Wait for the stepping pulse

Move to the sample

Lower the lift to the work position

Wait for the stepping pulse

Switch on the peristaltic pump for the sample transfer

Wait for the minimum pump time to be completed

Wait for the stepping pulse

Switch off the peristaltic pump