6.3 Lift settings (Lift)

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

863 Compact Autosampler

Files

Exported methods will be stored in this directory.
The directory will be created the first time a
method is exported.

Only methods being located in this directory can
be imported.

6.3

Lift settings (Lift)

Menu

System Lift

Work position

The working height of the lift can be set to the desired value. This is
accomplished by means of the direct operation of the lift.

Three functions can be selected from the function bar with [

] and []

and then executed by pressing [OK]:

Work pos. moves the lift to the current working height.

Up moves the lift 6 mm upward.

Down moves the lift 6 mm downward.

When this dialog page is exited, the respective current lift position will be
applied as Work position.

Range

0 ... 126 mm (Increment: 6)

Default value

60 mm

The lift can only be moved by steps of 6 mm. If more a accurate setting is
required, the needle position can be fine aligned.

Adjusting the needle position
First move the lift to working height.

1 Open the safety shield

Loosen the two lower fastening screws of the safety shield using the
hexagon key provided, and tip up the safety shield.

2 Loosen the nut

Slightly loosen the nut below the lift head using a wrench.