9 Troubleshooting

863 Compact Autosampler

■■■■■■■■

59

9 Troubleshooting

9.1

Problems and their solutions

Problem

Cause

Remedy

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set the contact pressure (see "Set the
contact pressure", page 17)
.

Peristaltic pump – pump
tubing defective.

Replace the pump tubing (see Figure 7, page
15)
.