4.2 Sample no air gap

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

863 Compact Autosampler

Lower the lift to the work position

Switch on the peristaltic pump for the sample transfer

Wait for the pump time to be completed

Switch off the peristaltic pump

Move the lift upward

Switch on the peristaltic pump

Wait for the waiting time for air gap to be completed

Switch off the peristaltic pump

Wait for the stepping pulse

4.2

Sample no air gap

This method template is suitable for simple determinations.

No air gap is formed in the transfer tubing between the individual sam-
ples.

Preparing and starting the sample series

1 Press [START].

2 Under Number of samples, enter the number of samples or the

maximum value of 99.

3 Under Next sample pos., enter the rack position of the first sample

to be processed.

4 Press [BACK].

5 Start the sample series in the IC Net software.

Note

The method run will be automatically restarted for every sample via
remote connection (Autostart). A stepping pulse of the connected
instrument is expected at the end of the method run. The 863 Compact
Autosampler must be stopped manually by means of the button
[STOP] after the processing of the last sample.

The individual steps of the method:

Move to the sample

Lower the lift to the work position

Switch on the peristaltic pump for the sample transfer