beautypg.com

3 maintenance, 1 parts of the suppressor, 2 regenerating the suppressor – Metrohm 882 Compact IC plus – Anion User Manual

Page 84: Maintenance, Figure 36, Parts of the suppressor

background image

5.11 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

882 Compact IC plus – Anion

parable conditions, you should normally work with a freshly regenerated
suppressor unit.

Caution

The suppressor must never be switched over in a dry state, as there is a
risk of jamming. If the suppressor is in a dry state, it must be rinsed for
at least 5 minutes before it may be switched over.

Caution

In the case of reduced capacity or high backpressure, the suppressor
must be regenerated (see Chapter 5.11.3.2, page 74), cleaned (see
Chapter 5.11.3.3, page 76)
or replaced (see Chapter 5.11.3.4, page
78)
.

5.11.3

Maintenance

5.11.3.1

Parts of the suppressor

1

2

3

4

5

Figure 36

Parts of the suppressor

1

Union nut

2

Connecting piece (6.2832.010)

3

Rotor

4

Housing

5

Slot in the housing

5.11.3.2

Regenerating the suppressor

If the suppressor units are loaded for a longer period with certain heavy
metals (e.g. iron) or organic impurities, these can no longer be completely
removed with the standard regeneration solution. The capacity of the sup-
pressor units is consequently affected, which can result in reduced phos-