3.14 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
882 Compact IC plus – Anion
2 Connecting the eluent outlet capillary
Note
Depending on the instrument type, the suppressor is either con-
nected directly with the detector or, if available and used, with the
MCS.
■
Connect the output capillary labeled with out either with the
detector (see manual of the detector).
or
■
Connect the output capillary labeled with out with the input in of
the MCS, using a long PEEK pressure screw (6.2744.090).
3 Connecting the rinsing solution inlet capillary
■
With a short PEEK pressure screw (6.2744.070), fasten the end of
the inlet capillary labeled with rinsing solution to the pump tub-
ing connection of the pump tubing which carries the rinsing solu-
tion.
4 Connecting the rinsing the solution outlet capillary
■
Guide the other end of the outlet capillary labeled with waste
rins. into a sufficiently large waste container and fasten it there.
5 Connecting the regeneration solution inlet capillary
■
With a short PEEK pressure screw (6.2744.070), fasten the end of
the inlet capillary labeled with regenerant to the pump tubing
connection of the pump tubing which carries the regeneration
solution.
6 Connecting the regeneration solution outlet capillary
■
Guide the other end of the outlet capillary labeled with waste
reg. into a sufficiently large waste container and fasten it there.