Metrohm 882 Compact IC plus – Anion User Manual

Page 54

background image

3.14 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

882 Compact IC plus – Anion

2 Connecting the eluent outlet capillary

Note

Depending on the instrument type, the suppressor is either con-
nected directly with the detector or, if available and used, with the
MCS.

Connect the output capillary labeled with out either with the
detector (see manual of the detector).

or

Connect the output capillary labeled with out with the input in of
the MCS, using a long PEEK pressure screw (6.2744.090).

3 Connecting the rinsing solution inlet capillary

With a short PEEK pressure screw (6.2744.070), fasten the end of
the inlet capillary labeled with rinsing solution to the pump tub-
ing connection of the pump tubing which carries the rinsing solu-
tion.

4 Connecting the rinsing the solution outlet capillary

Guide the other end of the outlet capillary labeled with waste
rins.
into a sufficiently large waste container and fasten it there.

5 Connecting the regeneration solution inlet capillary

With a short PEEK pressure screw (6.2744.070), fasten the end of
the inlet capillary labeled with regenerant to the pump tubing
connection of the pump tubing which carries the regeneration
solution.

6 Connecting the regeneration solution outlet capillary

Guide the other end of the outlet capillary labeled with waste
reg.
into a sufficiently large waste container and fasten it there.