2 pump tubing connection with filter, Figure 35, Pump tubing connection – changing the filter – Metrohm 882 Compact IC plus – Anion User Manual
Page 82
![background image](/manuals/586428/82/background.png)
5.10 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
882 Compact IC plus – Anion
Table 2
Pump tubing
Order
number
Name
Material
Inner
diameter
Use
6.1826.020
Pump tubing (blue/
blue), 2-stopper
PVC (Tygon
ST)
1.65 mm
Pump tubing for online IC
instruments and automa-
tion in voltammetry
6.1826.310
Pump tubing LFL
(orange/green), 3-
stopper
PVC (Tygon)
0.38 mm
Pump tubing for bromate
determination using the
triiodide method.
6.1826.320
Pump tubing LFL
(orange/yellow), 3-
stopper
PVC (Tygon)
0.48 mm
For suppressor solutions,
acceptor solutions for
inline dialysis and for inline
ultrafiltration.
6.1826.330
Pump tubing LFL
(orange/white), 3-
stopper
PVC (Tygon)
0.64 mm
No special applications.
6.1826.340
Pump tubing LFL
(black/black), 3-stop-
per
PVC (Tygon)
0.76 mm
For the sample solution in
inline dialysis.
6.1826.360
Pump tubing LFL
(white/white), 3-stop-
per
PVC (Tygon)
1.02 mm
For sample transfer.
6.1826.380
Pump tubing LFL
(gray/gray), 3-stopper
PVC (Tygon)
1.25 mm
For inline sample dilution.
6.1826.390
Pump tubing LFL (yel-
low/yellow), 3-stop-
per
PVC (Tygon)
1.37 mm
For the sample solution in
inline ultrafiltration.
5.10.2.2
Pump tubing connection with filter
The 6.2821.130 filters (35-2) should be changed every 3 months, more
frequently at higher backpressure.
1
2
3
Figure 35
Pump tubing connection – Changing the filter