5 Handling and maintenance

882 Compact IC plus – Anion

■■■■■■■■

63

1

Outlet valve holder

2

Outlet valve
Order number: 6.2824.160

3

Inlet valve
Order number: 6.2824.170

4

Inlet valve holder

2 Cleaning undissected valve

Clean contaminated or blocked valves initially without dismantling
them completely.

Rinse the valve in eluent flow and counterflow direction using a
spray bottle filled with ultrapure water, RBS solution or acetone.

The rinsing effect is further increased through a short treatment
(lasting for a maximum of 20 s) in an ultrasonic bath.

Note

Longer lasting ultrasonic baths can damage the ruby ball of the
valve.

Only if this cleaning is useless, dismantle the valves separately and
clean the components.

3 Dismantling valve

Dismantle every valve separately.

Note

For dismantling the valve the tool for valve cartridges (6.2617.020)
is required.

Place the valve with the seal faced downwards above the recess
in the holder.

Push the valve components out of the valve housing using the
needle of the tool.