beautypg.com

16 guard column – Metrohm 882 Compact IC plus – Anion User Manual

Page 56

background image

3.16 Guard column

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

882 Compact IC plus – Anion

3.16

Guard column

The use of guard columns serves for protecting the separation columns
and increasing their service life considerably. The guard columns available
from Metrohm represent either actual guard columns or are so-called
guard column cartridges which are used together with a cartridge holder.
The installation of a guard column cartridge in the associated holder is
described in the leaflet of the guard columns.

Note

Information regarding which guard column is suitable for your separa-
tion column can be found in the Metrohm IC Column Program
(which is available from your Metrohm agent), the leaflet provided
along with your separation column, the product information on the
separation column at

http://www.metrohm.com

(product area Ion

Chromatography), or obtained directly from your agent.

Caution

New guard columns are filled with solution and are sealed on both
sides with stoppers or caps, respectively. Before using the guard col-
umn, you need to ensure that this solution is miscible with the eluents
used (observe manufacturer's data).

Note

The guard column may only be installed after the initial start-up (see
Chapter 4.1, page 50)
of the instrument. Until then, use the coupling
(6.2744.040) instead of the guard and separation column.

Note

Metrohm recommends, always to work with guard columns. These pro-
tect the separation column and can be exchanged regularly.