2 operation, 1 supply bottle, 2 aspiration filter – Metrohm 882 Compact IC plus – Anion User Manual
Page 66: 3 changing the eluent, 5 high pressure pump, 1 protection, Operation, High pressure pump, Protection
5.5 High pressure pump
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
882 Compact IC plus – Anion
5.4.2
Operation
5.4.2.1
Supply bottle
The supply bottle with the eluent must be connected as indicated in chap-
ter 3.8.1, page 24. This is above all important for eluents with volatile sol-
vents (e.g. acetone).
Moreover, condensation must also be prevented in the eluent bottle. Drop
formation can change the concentration ratio in the eluent.
In the case of very sensitive measurements, we recommend that the elu-
ent be stirred constantly with a magnetic stirrer (e. g. the 2.801.0010 with
6.2070.000).
5.4.2.2
Aspiration filter
To protect the IC system against foreign particles, we recommend aspirat-
ing the eluents via a 6.2821.090 aspiration filter (9-2). This aspiration filter
must be replaced should it show signs of yellow discoloration (but no later
than every 3 months).
5.4.2.3
Changing the eluent
When changing the eluent, it must be ensured that no precipitates can
occur. Solutions following one another in direct succession must therefore
be miscible. If the system has to be rinsed organically, several solvents
with rising or falling lipophilia must be used.
5.5
High pressure pump
5.5.1
Protection
Caution
The pump head is filled ex works with methanol/ultrapure water. It
must be ensured that the eluent used is freely miscible with the solvent
remaining in the pump head.
To protect the high pressure pump against foreign particles, we recom-
mend that the eluent undergoes a microfiltration (filter 0.45 µm) before
being aspirated via a 6.2821.090 aspiration filter (see Chapter 5.4.2.2,
page 56).
Salt crystals between the piston and seal cause abrasion particles which
can find their way into the eluent. These lead to contaminated valves, a
rise in pressure and in extreme cases scratched pistons. It is therefore
essential to ensure that no precipitates can occur (see Chapter 5.4.2.3,
page 56).