Figure 22, Install pump tubing connection with filter – Metrohm 882 Compact IC plus – Anion User Manual
Page 49
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
882 Compact IC plus – Anion
■■■■■■■■
39
2 Connecting the aspiration side
Place a 6.2744.034 tubing olive (21-2) on the aspiration side of the
pump tubing.
3 Connecting the pressure side
Note
Depending on the use of the peristaltic pump, on the pressure side
you can either connect:
■
Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 22, page 39) or
■
Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 23, page 40).
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.
Case A: 6.2744.180 pump tubing connection with filter:
1
2
3
Figure 22
Install pump tubing connection with filter
1
Union nut
2
Adapter
3
Tubing olive with filter holder
■
Slide union nut (22-1) onto the pump tubing.
■
Select a suitable adapter (22-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 1,
page 40).
■
Place the tubing olive with filter holder (22-3) onto the pump tub-
ing.
■
Screw the union nut (22-1) onto the tubing olive (22-3).
or
Case B: 6.2744.160 pump tubing connection without filter: