10 peristaltic pump, 1 operation, 2 maintenance – Metrohm 882 Compact IC plus – Anion User Manual
Page 81: 1 pump tubing, Operation, Maintenance
■■■■■■■■■■■■■■■■■■■■■■
5 Handling and maintenance
882 Compact IC plus – Anion
■■■■■■■■
71
5.10
Peristaltic pump
5.10.1
Operation
The pumping capacity of the peristaltic pump depends on the drive speed
(set via software), the contact pressure and, above all, the internal diame-
ter of the pump tubing. Depending on the application, different pump
tubings are used.
Caution
The service life of the pump tubings also depends on the contact pres-
sure. Therefore fully lift the tubing cartridges by loosening the snap-
action lever (21-10) on the right-hand side if the peristaltic pump is to
be turned off for a longer period. Once set, the contact pressure
remains unaffected.
Caution
The 6.1826.xxx pump tubings consist of PVC or PP and therefore must
not be used for rinsing with solutions containing acetone. In this case,
use other pump tubings or use another pump for rinsing.
5.10.2
Maintenance
5.10.2.1
Pump tubing
The pump tubing used in the peristaltic pump is a consumable whose
service life is restricted.
The LFL pump tubing with 3 stoppers is stretched in the tubing cartridge
in such a way that it comes to rest between two stoppers. This results in
two possible positions for the tubing cartridge. If the pump tubing should
exhibit clear signs of wear, then this can be stretched a second time, in
the respective alternate position.
Therefore replace the pump tubing periodically, or when used perma-
nently approx. every 4 weeks .
Selecting the pump tubing
The pump tubing differs in material, diameter and hence also pumping
capacity. Depending on the application, different pump tubings are used.
The following table provides information on the properties and use of the
pump tubing: