3 Installation

882 Compact IC plus – Anion

■■■■■■■■

29

1

Connection capillary
PEEK capillary, connects main piston and
auxiliary piston.

2

PEEK pressure screw, short
(6.2744.070)

3

Outlet valve holder

4

Pump head (6.2824.110)

5

Fastening screws
For fastening the pump head.

6

Inlet valve holder

7

Pump head inlet capillary
PEEK capillary at the input of the pump
head.

8

Pressure screw
For connecting a PEEK capillary to the cou-
pling (13-

9

).

9

Coupling
For the connection of the eluent path at the
input of the high pressure pump. Can be
ordered together with the pressure screw
(13-

8

) under the number (6.2744.230).

10 Deaerating capillary

For aspirating the eluent when deaerating
the high pressure pump (see Chapter 3.9.2,
page 30)
.

11 Purge valve

For deaerating the high pressure pump.
With rotary knob in the center and pressure
sensor.

12 Connection capillary

For connecting the inline filter (see Chapter
3.10, page 32)
.

13 Connection capillary

Connects the output of the pump head with
the purge valve.

Note

The eluent aspiration capillary is already installed in the newly delivered
instrument. The following installation instructions need not be carried
out at the time of initial installation.

Connecting inlet to the high pressure pump

5

1

2

3

4

Figure 14

High pressure pump – Connect inlet