beautypg.com

Metrohm 882 Compact IC plus – Anion User Manual

Page 62

background image

4.2 Conditioning

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

882 Compact IC plus – Anion

Select (or create) a suitable method.

2 Preparing the instrument

Ensure that the column is correctly mounted according to the
flow direction indicated on the label (arrow must point in the
direction of flow).

Ensure that the eluent aspiration tubing is immersed in the eluent
and that there is enough eluent in the eluent bottle.

Ensure that the aspiration tubings for the auxiliary solutions
(regeneration solution and rinsing solution) are immersed into the
respective solutions and that the is enough solution in the bottles.

3 Checking leak-tightness

In MagIC Net™, start the equilibration.

Check all capillaries and their connections from the high pressure
pump to the detector for signs of liquid escaping.
If eluent escapes anywhere, tighten the corresponding pressure
screw or loosen the connection, check the end of the capillary,
shorten it with a capillary cutter if necessary, and restore the con-
nection.

4 Conditioning the system

Rinse the system with eluent until the required stability of the base-
line is attained (normally 30 minutes).

During this time, step the MSM to the next position every 10
minutes.

The instrument is now ready for measuring samples.