13.5 Measuring with Delta mode

1. Normal mode, if necessary

Key (lights OFF)

2. Touch the reference plane (P1)



3. Delta mode, press MIN/MAX key to

(Press 3 times the key)
(MIN,MAX,DELTA)

4. Move the part or the instrument sideways to the point

(P2).
Delta is memorized. (Delta = maximum-minimum)







13.6 Measuring of squareness.

1. Use a lever dial indicator and clamp it into the

holder (A).

2. Adjust the lever dial indicator against the surface to be

measured. Set the lever dial indicator to zero.


3. Normal mode, if necessary press

Key (lights OFF)


4. Press

Key, to set the display
at zero.
5. Move the lever dial indicator along the surface to be

measured.

6. Read the lever dial indicator for squareness value

and digital display for the height displacement value.

P1

P2

A

0.000

MM

102.678

MM

0.000

MM

0.022

MM

0.000

MM

12.567

MM