13.3 Measuring diameters and centerlines.

The probe constant value should already have been
memorized : (see page 13)

1. Normal mode, if necessary press

key, lights OFF

2. Touch the reference plane (P1)

green light ON

3. Press PRESET key quickly (>1 sec.),

to set the display at zero.


Realise the probe.

4. Change to Diameter/Centerline mode by pressing

key, orange light blink.


5. Place the probe inside the bore to be measured and

touch a point (P2) which is well away from the

reversal point.

6. Move the part or the instrument sideways to the

probe. The reversal point is memorized.



7. Touch a point (P3) well away from reversal point.




8. Move the part or the instrument sideways to the

probe. The diameter is displayed as well
as the

symbol.



9. Displaying the centerline can be done in

2 ways :

1. by releasing the probe.

(no possibility to ask again measured diameter)

2. by pressing the key
(Yellow light ON)
The centerline is displayed, as well as the

symbol.