beautypg.com

Bowers Group Trimos V+ Series Height Gauges User Manual

Page 16

background image

TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland

Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 [email protected] http://www.trimos.ch

Page 16



13.4 Measuring with min and max mode.

1. Normal mode, if necessary press

Key (lights OFF)

2. Touch the reference plane (P1)

(green light ON)

3. Press PRESET key quickly (>1 sec.)

to set the display at zero.


key

4. Place the probe inside the groove to be measured

and touch a point (P2) which is well away from the

reversal point.

5. Press MIN/MAX key to MIN




6. Move the part or the instrument sideways to the

probe. The reversal point is memorized.

The first value is displayed.


7. Press PRESET key to set the reversal point at zero.

key

8. Press MIN/MAX key to MAX




9. Place the probe inside the groove to be measured

and touch a point (P3) which is well away from the

reversal point.

10. Move the part or the instrument sideways to the probe.

The reversal point is memorized.


The second value is displayed.


0.000

MM

24.456

M
M

0.000

MM

125.231

MM

MIN

124.043

MM

MIN

0.000

MM

MIN

0.320

MM

MAX

76.686

MM

MAX

76.993

MM

MAX

124

.043

76.993

P1

P2

P3