Calibrating the Bird 5011

The Bird 5011 is a highly accurate terminating power
sensor. Over time, the sensor’s “zero value” (reading
with no applied RF power) can drift, making all
readings inaccurate by this value. For example, if the
zero value is –2 µW, measuring a 5 mW signal will
give a reading of 4.998 mW, a 0.04% error. Measuring
a 50 µW signal will give a reading of 48 µW, a 4%
error. To keep this drift from causing a significant
error, rezero the sensor as necessary.

NOTE: For best results, make sure the sensor has
been connected to the Site Analyzer and the SA
turned on for at least 5 minutes. Do not apply RF
power to the sensor during calibration.

CAUTION

When using a Bird 5011, do not exceed 2 W average

or 125 W peak power for 5 µs. Doing so will render

the sensor inoperative.

Press from

Measure

Power Mode

Press to

begin the

calibration

Fwd

0.00 mWatts

Offset

12.00

FullScale

Measure Power Mode

Configuration Screen

Recall

Setups

0.00

dB

0.01 Watts

Calibrate

5011

This will take about 45
seconds to complete. Do
you wish to continue?

[Enter/Esc]