Chapter 3

Calibration

Calibration Accessories

Calibration Combination including:

!

One 50 ohm load

!

One Open standard

!

One Short standard

Test Cable (optional) – Phase-stable for reliable,
consistent results.

NOTE: If a test cable is used to connect the Bird
Site Analyzer to the system under test, attach the
standards to the end of the test cable during
calibration.

Calibrating

NOTE: For best results, calibrate the Site Analyzer
immediately before taking measurements.

Press key

from either

Measure

Match or

Fault

Location

mode

825.00

MHz

841.00

Calibration Mode

Calibration: Off

1.70

1.00

VSWR

Open

Short

Load

Cal Done