Flash ROM Upgrade

The Site Analyzer uses flash-programmable ROM. It
is not necessary to replace the EPROM to upgrade the
firmware.

NOTE: The update process will erase the memory
of the Site Analyzer, including all saved traces and
setups. Use the PC software to save all traces to a
PC before updating the flash ROM.

!

Install the FlashLoader program on a PC.

!

Turn off the Site Analyzer and disconnect the

external power supply.

!

Connect the PC to the Site Analyzer’s serial port

with the supplied serial cable.

!

Run the FlashLoader and click the Begin button.

The status will change to “Waiting to Connect”.

!

Connect the Site Analyzer to the power supply. The

green LED on the SA will turn on and the Flash

Updater will display “Connected.”

!

Select the data file (e.g. 04May2001.sre) and click

“Open”.

!

Select “Yes” to accept the warning and update the

flash, or “No” if you do not want to proceed.

!

The update process takes 3 to 5 minutes. After this

the Site Analyzer will beep several times, the green

power LED will turn off and the amber charging

LED will start blinking.