Connecting a sensor, Measure power mode – Bird Technologies SA-1700-P User Manual
Page 82
Bird Site Analyzer
72
Connecting a Sensor
The antenna test port is only used for testing
unpowered systems. For power measurement, an
external power sensor connected to the Remote Power
Sensor Port must be used.
Connect the Bird Site Analyzer to power sensors using
the built-in serial port, labeled “Remote Power Sensor”
(see “Connection Description” on page 7). A 9 pin serial
cable should be used. When a sensor is properly
connected to the Site Analyzer, the sensor status
message will change from “Waiting for Sensor” to
“Sensor Connected” and the warning message will be
replaced by the main display.
CAUTION
+22 dBm max. input
Do not apply RF power to Antenna Test Port. Exceed-
ing the maximum input will damage the Site Analyzer.
A sensor is
properly
connected
Fwd
0.0 Watts
Refl
600.00
Match
0.00 Watts
0.00 Loss(dB)
Measure Power Mode
PwrUnit
Watts
Sensor Connected
MatchUnit
Rtn Loss
Display
Fwd