Changing GX-2009 Instrument Parameters • 77

Instrument Information, and Power Off download
commands will be selectable. Click the Instrument
Information
download command to retrieve the instrument
information from the GX-2009. If you wish to download data
before making changes, click Complete Download instead
of Instrument Information.

5. Click the Set button to display the Set Window. Use the

GX-2009 Status Frame and the Gas/Sensor Frame to
change parameters stored in the GX-2009.

6. To change the serial number, station ID, or user ID stored in

the GX-2009, click the field you wish to change (e.g., user
ID), use the backspace key to remove the current entry, then
type the new information.

7. To change the datalogging trend interval time, click on the

down arrow in the Interval Trend Time Field and select the
desired interval time in seconds. The available choices are
10, 20, 30, 60, 180, 300, and 600 seconds.

8. To change the channel parameters, click the field you wish to

Figure 62: Set Window

Click to
upload
changes to
GX-2009