RKI Instruments GX-2009 Data Logger Management Program User Manual
Page 16
16 • Control Buttons
NOTE: If the GX-2009 Data Logger Management Program
database is on the user’s computer (default setting),
data will only appear for instruments downloaded to that
computer. If the database is on a network, data will
appear for all instruments downloaded to that database.
See “Accessing a Common Database” on page 20 for
more instruction.
The following types of data files are saved in the Data window:
•
Bump Test Files
A bump test file is saved for each instrument that has been
downloaded. It records the bump test information for every
bump test that was downloaded. The GX-2009 can save
information for up to 100 bump tests and calibrations
combined in its memory.
•
Calibration History Files
A calibration history file is saved for each instrument that has
been downloaded. It records the calibration information for
every calibration that was downloaded. The GX-2009 can
save information for up to 100 bump tests and calibrations in
its memory.
•
Interval Trend Data Files
Interval trend data is logged at the interval time defined in the
GX-2009. Each logged point is an average reading over the
previous time interval.
NOTE: Although the interval time is stored in the GX-2009, it is
not accessible using the GX-2009’s operational modes.
This parameter can only be accessed and updated
using either the Data Logger Management Program or
the User Setup Program.