beautypg.com

RKI Instruments GX-2009 Data Logger Management Program User Manual

Page 48

background image

48 • Viewing, Printing, Exporting, and Deleting Data in the Data Window

In table format, the interval trend data is shown as the
average gas readings over the user defined interval trend
time. So if the data logging session started at 4:13:38 PM
and the interval time is set to 1 minute, then the readings
logged at 4:14:38 PM are the average reading for each
channel over that one minute period.

Events are displayed on the screen under the channel in
which they occur and with the time of the event. Events
are displayed whether they occurred at scheduled log
times or in between them. Events include gas alarms such
as a warning condition, trouble conditions such as a low
sensor failure, and an indication that the unit is returning to
“normal” condition after an alarm has been reset using the

Figure 32: Interval Trend Data in Table Format

Click to see

Click to save data to a file

Click to show

Select table or

graph view

Click to see

Click to print data

condensed data

session summary

events only

Click to return to

Data Window