8 fault messages, Fault messages – Retsch AS 200 jet User Manual
Page 46

Fault messages
46
Pos : 15.2 /0020 Übersc hrift en/1. Übersc hrift en/1 Fehl ermeldungen @ 2 \ mod_1259056589728_9. doc x @ 17580 @ 1 @ 1
8 Fault messages
Pos : 15.3 /0005 RET SCH/ 0005 R ETSCH Bedienungsanlei tungen Kapit elsamml ung en/AS200Jet/ 0013 AS200J et Fehler meldung en/1301 AS200 Jet Modul F ehl ermel dungen @ 2\ mod_1263548251088_9.doc x @ 18608 @ @ 1
E11 FAULT DRIVE/MOTOR
Service necessary
E20 FAULT CONTROLLER
Service necessary
E24 FAULT VALVE
Connection to electrical vacuum force adjustment
disconnected.
•
Confirm the message in the display (C) by
pressing the operating button (F) on the
operating element.
E83 VACUUM TOO LOW
Check whether
•
The suction apparatus is connected;
•
The suction apparatus is generating
sufficient vacuum;
•
The collecting receptacle in the suction
apparatus is full;
•
The sieve cover is attached.
E84 DROP IN VACUUM
Check whether
•
The suction apparatus is connected;
•
The suction apparatus is generating
sufficient vacuum;
•
The collecting receptacle in the suction
apparatus is full;
•
The sieve cover is attached.
H45 INTERRUPTION DUE TO MAINS FAILURE Restart the device