Transport, scope of delivery, installation, Warning – Retsch AS 200 jet User Manual
Page 17

Transport, scope of delivery, installation
17
•
Use the supplied connection cable to connect the device to the mains power
supply.
Pos : 8. 18 /0005 RET SCH/ 0005 R ETSCH Bedienungsanlei tungen Kapit elsamml ung en/AS200Jet/ 0006 AS200J et Transport, Li ef erumf ang und Auf st ellen/AS200J et M odul ext erne Absicherung @ 1 \ mod_1243926769314_9. doc x @ 10385 @ @ 1
The external fuse must be at least T15A (230V) T15A (100/120V).
Pos : 8. 19 /0025 War nhi nweis e/ W0015 WARNUNG Kabelbr uc h Stroms toß Er dung @ 4 \mod_1334064762171_9. doc x @ 27560 @ @ 1
WARNING
W0002
Danger to life through electric shock
–
An electric shock can lead to burns and to cardiac arrhythmias or to
respiratory arrest and cardiac arrest.
•
The device may only be operated with plugs that have a protective
conductor (earthed).
Pos : 9. 1 / 0005 R ETSCH /0005 RET SCH Bedienungs anl eit ung en Kapi tels ammlungen/------- Seit enumbruc h ----------- @ 0\ mod_1222344373758_0. doc x @ 2386 @ @ 1